Energy spectra, angular spread and dose distributions of electron beams
from various accelerators used in radiotherapy
Next: Contents
Use the Up button to get back to this page from within the document.
A pdf version of the entire report is available
here
©
Copyright 1995, National Research Council of Canada Ottawa
Energy spectra, angular spread and dose distributions of electron beams
from various accelerators used in radiotherapy
PIRS-0439
G.X. Ding and D.W.O. Rogers
Institute for
National Measurement Standards
National
Research Council of Canada
Ottawa, Canada K1A 0R6
TEL:
(613) 993-2715
FAX: (613) 952-9865
E-mail:
gding@irs.phy.nrc.ca
or
E-mail: dave@irs.phy.nrc.ca (drogers at
physics.carleton.ca since Dec 2003)
ABSTRACT
Using the EGS4 Monte Carlo code, BEAM, we have simulated electron beams
used in radiotherapy.
The simulation covers
electron beams from various accelerators,
including the NRC research accelerator,
a Varian Clinac 2100C, a Philips SL75-20, a Siemens KD2,
an AECL Therac 20, and a Scanditronix MM50.
The simulated realistic beams are stored in phase-space data files.
A phase-space data file contains details of each particle's information
at the phantom surface. The phase-space files are used as inputs to
the EGS4 code DOSXYZ to calculate dose distributions in water phantoms.
The phase-space files are also analyzed to obtain the
characteristics of the electron beams.
This report systematically presents detailed information
and characteristics of up to 5 electron beams from each of
these medical accelerators.
The accuracy of the simulated clinical beams is demonstrated by
comparing the calculated dose distributions
with measured dose distributions in water phantoms.
These beams with an energy range of 5 - 50 MeV
are commonly used for radiotherapy in cancer clinics.
The comprehensive beam data presented here
enhance our knowledge on many aspects of clinical electron beams.
- Contents
- List of Tables
- List of Figures
- Introduction
- Monte Carlo Calculation
- Results and Discussion
- Conclusion
- Acknowledgments
- Tables
- Figures: Digital Data linked to UL etc in headings
- Figure 1: NRC accelerator view
- Figure 2: Clinac 2100C accelerator view: type II applicator
- Figure 3: Clinac 2100C accelerator view: type III applicator
- Figure 4: Clinac 2100C accelerator view: modified type II applicator
- Figure 5: Philips SL75-20 accelerator view
- Figure 6: Siemens KD2 accelerator view
- Figure 7: Therac 20 accelerator view
- Figure 8: Racetrack MM50 accelerator view
- Figure 9: NRC 10 MeV beam data. Digital: UL, UR, L
- Figure 10: NRC 20 MeV beam data. Digital: UL, UR, LL, LR
- Figure 11: Clinac 2100C 6-MeV beam data. Digital: UL, UR, L
- Figure 12: Clinac 2100C 9-MeV beam data. Digital: UL, UR, LL, LR
- Figure 13: Clinac 2100C 12-MeV beam data. Digital: UL, UR, L
- Figure 14: Clinac 2100C 15-MeV beam data. Digital: UL, UR, L
- Figure 15: Clinac 2100C 18-MeV beam data. Digital: UL, UR, LL, LR
- Figure 16: Clinac 2100C with Type III applicator 9-MeV beam data. Digital: UL, UR, LL, LR
- Figure 17: Clinac 2100C: modified Type II appl, 9-MeV beam data. Digital: UL, UR, LL, LR
- Figure 18: Clinac 2100C 9-MeV profiles - 3 applicators. Digital: UL, UR, L
- Figure 19: Philips SL75-20 5-MeV beam data. Digital: UL, UR, L
- Figure 20: Philips SL75-20 10-MeV beam data. Digital: UL, UR, L
- Figure 21: Philips SL75-20 14-MeV beam data. Digital: UL, UR, L
- Figure 22: Philips SL75-20 17-MeV beam data. Digital: UL, UR, L
- Figure 23: Philips SL75-20 20-MeV beam data. Digital: UL, UR, L
- Figure 24: Philips SL75-20 5-MeV profiles. Digital: U, L
- Figure 25: Philips SL75-20 20-MeV double peak. Digital: L
- Figure 26: Siemens KD2 6-MeV beam data. Digital: UL, UR, L
- Figure 27: Siemens KD2 11-MeV beam data. Digital: UL, UR, L
- Figure 28: Siemens KD2 21-MeV beam data. Digital: UL, UR, L
- Figure 29: Siemens KD2 11-MeV profiles. Digital: U, L
- Figure 30: AECL Therac 20 6-MeV profiles. Digital: UL, UR, L
- Figure 31: AECL Therac 20 6-MeV profiles. Digital: U, L
- Figure 32: AECL Therac 20 9-MeV beam data. Digital: UL, UR, L
- Figure 33: AECL Therac 20 13-MeV beam data. Digital: UL, UR, L
- Figure 34: AECL Therac 20 17-MeV beam data. Digital: UL, UR, L
- Figure 35: AECL Therac 20 20-MeV beam data. Digital: UL, UR, L
- Figure 36: Racetrack MM50 25-MeV beam data. Digital: UL, UR, L
- Figure 37: Racetrack MM50 40-MeV beam data. Digital: UL, UR, L
- Figure 38: Racetrack MM50 50-MeV beam data. Digital: UL, UR, L
- Figure 39: Racetrack MM50 dose profiles. Digital: U, L
- References
- About this document ...