A fourth analysis of 2D GEM position resolution

Dean Karlen / November 14, 2000

This document summarizes a quick analysis of the GEM data taken from November 13, 2000. The analysis follows the general approaches described in the third analysis. The goal of this analysis is to check the data for any serious problems before a careful set of data is taken for publication. The gas is P10 and the preamps were ALEPH. The scope time scale was decreased by a factor of 2 compared to the October data... the data recorded here therefore correspond to 250 MHz sampling.
 

Index


 
 

GEM pad layout

The figure linked here shows the assumed GEM layout and coordinate system. Dimensions are in mm. The pads are numbered from 1 to 8, according to the readout channel. The central pad is read out by both oscilloscopes (channels 1 and 5), to provide a common trigger. The coordinates of a few points are shown in colour.

GEM data

The data sets taken on November 13 were taken with P10 gas with Vdrift = 3475V, Vgem=3375, using the ALEPH preamplifier, and the xray tube set at 6 kV. About 400 events are taken for each run.

The data runs are summarized below. The collimator location is indicated using the coordinate system described above.
 
 
run number  x_coll (mm)  y_coll (mm)
1101  0.  1.443 
1102  0.  1.643 
1103  0.  1.243 
1104  0.  1.043 
1105  0.  0.843 

GEM data analysis programs

The results shown below come from the gemanal program (version 0.8) located in the directory /home/karlen/gem. An associated paw kumac file, gemanal.kumac, is found in the same area.

Gain variation

The gain of the system was not constant over these runs, as can be seen in the figure linked here. There was quite a large change from run 1103 to 1104.

Pedestals

The data for each channel is corrected by using a pedestal defined by the average of measurements before the pulse (time bins 10-60).

The figure linked here shows data from a typical event, after pedestal correction, when the x-ray collimator was positioned over the coordinate (0.,1.243) (mm).

Some plots have been produced to check if the baseline changes from run to run: The pulse shape is fit to a quadratic 300ns - 900 ns after the induced pulse. The value of the fit at the point 600 ns after the pulse defines the "baseline". The value of the baseline for the different runs are shown in the figure linked here. The baseline in pads 4 and 6 are roughly constant. The baseline in pads 3 and 7 reduce smoothly as the collimator moves closer to the centre of pad 1. Unlike the 2nd analysis, the variation is consistent with a small amount of charge sharing: pads 4 and 7 behave similarly as do pads 3 and 6.

Separation of direct and induced components of signals

As in the 2nd analysis, the direct charge component of a signal is deduced from the amplitude measured a fixed time after the peak. In this analysis, the delay is chosen to be 1200 ns (compared with 300 ns in the 2nd analysis).

The figure linked here shows the mean ratio of the "late" to the peak amplitudes on pads 1,2, and 8 for different collimator positions. The error bars indicate the standard deviations of the ratio. (The mean and standard deviations are found by fitting each ratio distribution to a Gaussian). Since the standard deviation is less than 1%, the late amplitude is used instead of the peak amplitude for the charge fraction determination of both direct and mixed signals. The ratios differ for pads 1, 2, and 8, by only about 2%.

Position analysis from direct charge sharing

Determination of charge
The data from t The data from these runs were used to map out the pad response function for direct charge collection. The charge collected by a pad is assumed to be proportional to the peak amplitude of the pulse for signals dominated by direct charge collection. Rather than use the peak amplitude (VP) directly, the "late" amplitude (AF) is used, and scaled to a new value (AN) corresponding to the peak amplitude,  as follows:
AN_pad = AF_pad / R_pad
where R_pad is the mean ratio of late to peak amplitudes, from the previous section. R_1 = 0.707, R_2 = 0.721, and R_8 = 0.723.
Observed charge fraction in pad 1 - determination of cloud size
The figure linked here shows the observed charge fraction in pad 1, as a function of the y-coordinate of the collimator. The solid curve shows the prediction of the model when sigma_x=0.56 mm. The dashed curves show the predictions for sigma_x = 0.53 and 0.59 mm. There appears to be a problem with the last two runs. The charge fractions in pad 1 are not consistent with expectations. Probably, the location of the x-ray collimator is not recorded correctly. Note that these same two runs also have a much smaller gain. In the future, these sorts of checks should be done during data taking.
Determining position from charge fractions
The figure linked here shows histograms of the x and y coordinate estimates, with respect to the x and y collimator position, for run 1101. Fitting the distributions to Gaussians, gives central values of (-0.010 mm,0.000 mm) and standard deviations of 62 and 59 microns in x and y, respectively.

Position analysis from induced pulses

To determine the coordinate from induced pulses, the same approach is used as in the 2nd analysis. The amplitude of the induced pulse is assumed to be proportional to the total charge of the event and a function of the distance from the cloud centroid to the pad centre. An algorithm combines the radial information from all pads that have an induced pulse to determine the x-ray location.

The data taken along the line (x=0) is used to characterize the induced response function. The ratio of the peak amplitude of the induced pulse to the total charge of the event is shown as a function of distance to the pad centre is shown in the figure linked here. The response function from the various pads do not line up. Simple scaling by gain factors will not improve the agreement. This further indicates that a problem exists with the data... the location of the xray collimator is possibly incorrect.

Conclusion

The data from November 13, 2000 shows some problems: the x-ray location and gain of the two final runs are anomalous.

Update (November 14)

By looking at the individual channels for the different runs, it is evident that the problem that occurs for the last two runs is most pronounced in the centre pad (channels 1 & 5). It appears that the pre-amp gain for that pad was reduced by a factor of 1.7. When correcting for this, the gain variation is much more stable, the variation in the charge fraction in pad 1 follows the expected form with sigma_x = 600 microns, and the ratio of the peak amplitude of the induced pulse to the total charge of the event lines up much better. The connection for pad 1 and the preamp gain for the pad should be checked.

Update (November 20)

Source of problem was identified: Incorrect termination of signals to scope for channels 1 & 5.
© 2006 Carleton University 1125 Colonel By Drive, Ottawa, Ontario, K1S 5B6 Canada (613) 520-7400
| Contacts |
Canada's Capital University