The data runs are summarized below. The collimator location is indicated using the coordinate system described above(ie. with respect to x=0 located on edge between channels 3 and 5). The first digit of the run number is the day of the month that the data was taken. Non-coincident triggers of the scopes resulted in the loss of only 10-20 events per run for most runs (much improved from analysis 6 data). The second last column indicates the GEM-strip number (1-10) that corresponds to the strip number 1 in this analysis, and the last column indicates the GEM-strip number for those events where the pinhole was centred on a strip.
As an example, an event from run 104 is shown here. The far away strips have very small induced pulses. There is less noise than in the analysis 6 data. The problematic crosstalk (order 5-10mV) seen in distant strips (and presumably present on all strips) is still present in the data. The amplitude of these signals is correlated with the total charge signal observed. It is interesting to note that the amplitude is approximately the same in all strips, independant of its distance from the signal strip. When the pinhole is centred over a strip, a small amount of charge is collected by the neighbouring strips, that partially cancels out the positive crosstalk signal.
|
|
|
|
GEM-strip centre |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
The scaling factor for all pads is taken to be 0.44 for runs 101-205 and 0.71 for runs 601-704.
The ratio of the peak amplitude of the induced pulse to the total charge of the event as a function of distance to the strip centre is shown in the figure linked here. No modification of the strip gains were applied to bring the measurements into agreement. In analysis 6, corrections of 5-8% were necessary.